11-34
Triggering
2400 Series SourceMeter
®
User’s Manual
Output triggers
All three output triggers are available in the trigger layer for the Pulse Mode. If enabled,
the output triggers occur where indicated in the trigger model.
Invalid trigger settings
Front panel operation
For front panel operation, most trigger settings are done from the “CONFIGURE TRIG-
GER” menu which is accessed by pressing CONFIG and then TRIG. For Pulse Mode
operation, the following settings for the “TRIG LAYER” are ignored.
DELAY — Trigger delay is not used. Therefore, the trigger DELAY setting is ignored.
TRIGGER IN EVENTS — The source and delay event detectors are not used in the
Pulse Mode. Therefore, the SOURCE and DELAY settings (on/off) are ignored.
TRIGGER OUT EVENTS — For the DC Mode, source delay is set from the source con-
figuration menu, which is accessed by pressing CONFIG and then SOURCE V or I. When
the Pulse Mode is selected, the source delay menu items (DELAY and AUTO DELAY) are
replaced by the pulse DELAY and PULSE WIDTH menu items.
Remote operation
The following commands are not valid in the Pulse Mode and are therefore ignored:
:TRIGger:DELay — Trigger delay is not used in the Pulse Mode.
:SOURce:DELay and :SOURce:DELay:AUTO — Source delay is not used in the Pulse
Mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...