18-76
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
2425 and 2430 DC Mode
<n> =
-3.15 to 3.15
Specify I-Source level (amps)
-105 to 105
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
20V range (V-Source)
MINimum
10µA range (I-Source)
200mV range (V-Source)
MAXimum
3A range (I-Source)
100V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
2430 Pulse Mode
<n> =
-10.5 to 10.5
Specify I-Source level (amps)
-105 to 105
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
20V range (V-Source)
MINimum
10µA range (I-Source)
200mV range (V-Source)
MAXimum
10A range (I-Source)
100V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
2440
<n> =
-5.25 to 5.25
Specify I-Source level (amps)
-42 to 42
Specify V-Source level (volts)
DEFault
100µA range (I-Source)
10V range (V-Source)
MINimum
10µA range (I-Source)
200mV range (V-Source)
MAXimum
5A range (I-Source)
40V range (V-Source)
UP
Select next higher range
DOWN
Select next lower range
Query
:RANGe?
Query range for specified source
:RANGe? DEFault
Query *RST default source range
:RANGe? MINimum
Query lowest source range
:RANGe? MAXimum
Query highest source range
Description
This command is used to manually select the range for the specified
source. Range is selected by specifying the approximate source magni-
tude that you will be using. The instrument will then go to the lowest
range that can accommodate that level. For example, if you expect to
source levels around 3V, send the following command:
:SOURce:VOLTage:RANGe 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for SourceMeter 2400
Page 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...