background image

  

 

Ultra

Max

O2

O P E R A T I N G   M A N U A L   & 

I N S T R U C T I O N S   F O R   U S E

R 2 2 1 P 1 1

R221M11 Rev. C

O X Y G E N   A N A LY Z E R

Summary of Contents for UltraMaxO2

Page 1: ...UltraMaxO2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 2 1 P 1 1 R221M11 Rev C O X Y G E N A N A LY Z E R...

Page 2: ......

Page 3: ...ieu of all other warranties expressed or implied including warranty of merchantability and fitness for a particular purpose CLASSIFICATIONS Classification Class II electrical device Protection against...

Page 4: ...ther than conventional oxygen concentrators Not for use with inhalation agents Operating the UltraMaxO2 in flammable or explosive environments may result in fire or explosion Not suitable for use in t...

Page 5: ...in minor or moderate injury and property damage Replace the batteries with high quality AA Alkaline or Lithium batteries DO NOT use rechargeable batteries When not in use for periods greater than 30...

Page 6: ...1 Oxygen Flow and Pressure Measurement 3 2 2 Switching Pressure Units of Measure 3 2 3 Calibration Verification Procedure 3 3 0 FACTORS INFLUENCING ACCURATE READINGS 4 3 1 Effects of Temperature 4 3...

Page 7: ...following features and operational benefits Accurate oxygen measurements No in field calibration required Convenient ability to measure pressure in PSI or kPa Durable compact design Large easy to rea...

Page 8: ...nated next to the flow measurement Not shown when in calibration verification mode KPA Indicates the pressure measurement is in units of kilopascals PSI Indicates the pressure measurement is in units...

Page 9: ...pressure cover the gas sample outlet with thumb or finger while gas is flowing 7 Wait 5 seconds for the display to read pressure 2 2 Switching Pressure Units of Measure The UltraMaxO2 can measure pre...

Page 10: ...gas from air leaving concentrated oxygen and argon at a specific oxygen to argon ratio This operating principle may be altered when concentrators are set to flow at the low end of their operational r...

Page 11: ...closer to room temperature before use E08 Device self check found error Recommended Action Remove and replace the batteries If error code repeats return the UltraMaxO2 to the manufacturer for factory...

Page 12: ...ent Range 0 10 LPM Flow Measurement Accuracy 0 2 LPM Flow Measurement Resolution 0 1 LPM PRESSURE PSI kPa Pressure Measurement Range 0 5 50 3 4 344 Pressure Measurement Accuracy 0 5 0 5 Pressure Measu...

Page 13: ...Your Unit Part Number Item R211M11 Operating Manual and Instructions for Use RP46P11 003 Gas Sample Tubing 8 2 Optional Accessories Part Number Item R221P15 Soft Cover An electronic version of this Op...