88
Timing Chart
Timing chart for the continuous test mode
This is the timing chart for measurement performed by inputting the START and STOP signals from
the EXT. I/O.
Settings
Test time (TIMER)
OFF or ON
Comparator delay (COMP DELAY)
MANUAL 0.001 s
Test mode (COMP MODE)
CONTINUE
Contact check (CONTACT CHECK)
OFF
BDD
OFF
BDD stop function (BDD STOP)
OFF
START
(Input)
STOP
(Input)
TEST
(Output)
VON
(Output)
Test voltage
Measurement
status
PASS
UPPER FAIL
LOWER FAIL
(Output)
Screen display
(Measured
value)
t2
Set voltage
High-voltage application start
The voltage monitor value is 90% or
more of the set voltage.
(While a voltage is applied)
Stop
Measurement-1
Previous result
The test automatically ends when PASS STOP or
FAIL STOP is set.
----
OFF
Discharge start
Discharge
complete
Test end
Discharge
Measurement
delay
Measurement-2 Measurement-3
Measurement-N
Stop
Measurement
Previous result
Measurement-2
result
Measurement-1
result
Measurement-3
result
Measurement-N result
Measurement-1 Measurement-2 Measurement-3
Measurement-N
Description
Time
t1
Pulse width of the START or STOP signal
3 ms min.
t2
Detection time of the START or STOP signal
4 ms max.
t3
Test time
Based on the setting
t4
Measurement time
50 Hz: 20 ms
×
PLC setting
60 Hz: 16.7 ms
×
PLC setting
t5
Calculation time
1 ms max.
t6
Discharge time
20 ms max. (at pure resistance measurement)
t7
START setup time
40 ms
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...