146
Communications Command Reference
Querying the contact check capacitance value
Syntax
Query
:CONTactcheck:CAPacitance?
Response
(9 bytes +
CRLF)
<Capacitance value>
•
Capacitance value {0.0E-09 to 200.0E-09|999.9E-09}
Returns the capacitance value.
Description
Queries the capacitance value when a contact check is performed.
When the capacitance value is 200 nF or more (OVER), 999.9E-09 is returned.
Example
:CONTactcheck:CAPacitance?
␣
20.0E-09
The capacitance value for the contact check is 20 nF.
Querying the contact check judgment result
Syntax
Query
:CONTactcheck:RESult?
Response
(4 bytes +
CRLF)
<Contact check judgment result>
•
Contact check judgment result {NONE|PASS|FAIL}
Returns the judgment result of the contact check.
Description
Returns the judgment result of the contact check.
NONE
The contact check function is disabled, or there is no judgment result (not
measured).
PASS
The judgment result of the contact check is PASS.
When the capacitance value is more than 200 nF (OVER), PASS is
returned as well.
FAIL
The judgment result of the contact check is FAIL.
Example
:CONTactcheck:RESult?
FAIL
The judgment result of the contact check is FAIL.
BDD
Setting the BDD judgment based on the CC (charging) voltage
Syntax
Command
:BDD:CC:V {ON|OFF}
Description
Sets the BDD judgment using the CC (charging) voltage.
Example
:BDD:CC:V ON
The BDD judgment is set to be performed using the CC (charging) voltage.
Querying the BDD judgment using the CC (charging) voltage
Syntax
Query
:BDD:CC:V?
Response
(2 or 3 bytes +
CRLF)
{ON|OFF}
Description
Queries the BDD judgment setting using the CC (charging) voltage.
Example
:BDD:CC:V?
ON
The BDD judgment using the CC (charging) voltage is set.
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...