36
Setting the Range (Auto/Manual)
3.2 Setting the Range (Auto/Manual)
There are two range settings, auto range and manual range. For the manual range, you can select
from four settings, 2 M
Ω
, 20 M
Ω
, 200 M
Ω
, and 2000 M
Ω
. The range that can be set, however,
varies depending on the test voltage.
See “9.2 Input, Output, and Measurement Specifications” (p. 180).
In the auto range, range movement takes place after the test starts. It may take time for the
measured value to be displayed. In order to eliminate the wait time, set the range to manual
according to the DUT.
Setting the test voltage in the Measurement screen
1
Set the range to the edit mode.
[RANGE]
Select
2
Change and confirm the range.
Select the range
[AUTO]
Set the auto range
[HOLD]
Set the manual range
[ ↓ ]
Lower the range
[ ↑ ]
Raise the range
Confirm
[CANCEL]
Cancel
HIOKI BT5525A961-00