44
Judging the Measured Value (Comparator Function)
3.6 Judging the Measured Value
(Comparator Function)
By using the comparator function, the following can be performed.
Displaying the judgment result on the instrument (COMP lamp U. FAIL / PASS / L. FAIL)
U. FAIL
Measured value > Upper limit value
PASS
Lower limit value ≤ Measured value ≤ Upper limit value (within the judgment reference)
L. FAIL
Measured value < Lower limit value
Sounding a buzzer (Default: FAIL)
See “3.8 Notifying the Judgment Result and Test Completion with a Beep” (p. 47).
Externally outputting the judgment result
See “7 External Control (EXT. I/O)” (p. 83).
There are the following three judgment methods. The setting range of the upper/lower limit values
is from 0.000 M
Ω
to 9999 M
Ω
.
Using the lower limit
value for a judgment
Example: 100 M
Ω
(lower limit value)
PASS
LOWER FAIL (L. FAIL)
Lower limit
value (M
Ω
)
Using the upper limit
value for a judgment
Example: 100 M
Ω
(upper limit value)
PASS
UPPER FAIL (U. FAIL)
Upper limit
value (M
Ω
)
Using the upper and
lower limit values for
a judgment
Example: 1000 M
Ω
(upper limit value)
Example: 100 M
Ω
(lower limit value)
Lower limit
value (M
Ω
)
LOWER FAIL (L. FAIL)
PASS
UPPER FAIL (U. FAIL)
Upper limit
value (M
Ω
)
IMPORTANT
In the following circumstances,
[U. FAIL]
and
[L. FAIL]
light up simultaneously (UL.FAIL):
• When no judgment is possible because the measured value is
[Over.F]
(overflow) due to the
upper or lower limit value being greater than the display range’s upper limit.
• When no judgment is possible because the measured value is
[Under.F]
(underflow) due to the
upper or lower limit value being less than the display range’s lower limit.
• When the BDD count is 1 or greater while the comparator function and BDD judgment are
enabled.
See “Enabling BDD judgment” (p. 63).
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...