122
Communications Method
Standard Event Status Enable Register (SESER)
When each bit of the Standard Event Status Enable Register (SESER) is set to “1”, the ESB bit of
the Status Byte Register (STB) is set to “1” if any corresponding bit of the Standard Event Status
Register (SESR) changes from “0” to “1” and the ESB bit is set to “0” when all the bits become “0”.
Status Byte Register (STB)
Bit 6
Bit 5
Bit 4
SRQ
MSS
ESB
MAV
Standard Event Status Register (SESR)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
PON
(URQ)
CME
EXE
DDE
QYE
(RQC)
OPC
AND
AND
AND
AND
AND
OR
AND
PON
(URQ)
CME
EXE
DDE
QYE
(RQC)
OPC
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Standard Event Status Enable Register (SESER)
Queries and settings for registers
Register
Query
Settings
Status Byte Register
*STB?
–
Service Request Enable Register
*SRE?
*SRE
Standard Event Status Register
*ESR?
–
Standard Event Status Enable Register
*ESE?
*ESE
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...