Limiting the Current to Be Applied to the Device Under Test (DUT)
5.3 Limiting the Current to Be Applied to the Device
Under Test (DUT)
The current to be applied to the DUT can be limited (current limit function).
50 μA to 0.99 mA
10 μA resolution
1.0 mA to 50.0 mA
100 μA resolution
The charging time for the capacitive load can be shortened. It is also possible to reduce the risk
which the DUT can be damaged or destroyed.
The current limit value can be set manually or automatically.
When setting the current limit value automatically, it will be automatically calculated based on the
test voltage, the charging time, and the DUT’s electrostatic capacitance value.
Calculating formula
Current limit value
(
A
)
=
Electrostatic c
apacitance value
(
F
)
× Test voltage
(
V
)
Charging time
(
s
)
The DUT’s electrostatic capacitance value can be either set manually (MANUAL) or measured and
set automatically at the start of testing (AUTO).
Current
limit
CURRENT
LIMIT
setting
CHARGE
TIME
setting
DUT
CAPACITY
setting
Operation
Applications
Manual
MANUAL,
current
limit value
–
–
Limits the current applied to
the current value set with the
CURRENT LIMIT setting.
Use when you wish to
limit the applied current to
a user-specified value.
Auto
AUTO
Charging
time
MANUAL,
electrostatic
capacitance
value
Limits the current applied to
the current value calculated
based on the test voltage,
the charging time (as set
by CHARGE TIME), and
the DUT’s electrostatic
capacitance value (as set by
DUT CAPACITY).
Use when you wish to
limit the applied current to
a user-specified charging
time for a DUT whose
electrostatic capacitance
value is known.
AUTO
Limits the current applied to
the current value calculated
based on the test voltage,
the charging time (as set by
CHARGE TIME), and the
electrostatic capacitance
value as measured at the
start of the test.
Since electrostatic
capacitance measurement
is the same type of
measurement as performed
by the contact check function,
the test time will increase by
the electrostatic capacitance
measurement time. In
addition, the contact check
signal will be applied to the
DUT at the start of the test.
Use when you wish to
limit the applied current to
a user-specified charging
time for a DUT whose
electrostatic capacitance
value is not known.
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...