87
Timing Chart
Panel load signal correspondence table
LOAD0 to LOAD3
Panel number
LOAD3
LOAD2
LOAD1
LOAD0
No load
OFF
OFF
OFF
OFF
1
OFF
OFF
OFF
ON
2
OFF
OFF
ON
OFF
3
OFF
OFF
ON
ON
4
OFF
ON
OFF
OFF
5
OFF
ON
OFF
ON
6
OFF
ON
ON
OFF
7
OFF
ON
ON
ON
8
ON
OFF
OFF
OFF
9
ON
OFF
OFF
ON
10
ON
OFF
ON
OFF
11
ON
OFF
ON
ON
12
ON
ON
OFF
OFF
13
ON
ON
OFF
ON
14
ON
ON
ON
OFF
15
ON
ON
ON
ON
ON/OFF in the above table indicates the status of the external switch (SW) or transistor.
7.2 Timing Chart
The level of each signal indicates the ON/OFF status of the contact. When using the current source
(PNP) setting, the voltage level is High when contact is ON and Low when contact is OFF.
For the voltage level in the current sink (NPN) setting, High and Low are reversed.
IMPORTANT
In the following cases, up to 1 s is added to the START signal detection time.
• The START signal is input after the test voltage is changed.
•
The test voltage is changed using the LOAD signal or communications command.
External Control (EXT
. I/O)
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...