135
Communications Command Reference
Error
Reading and clearing errors
Syntax
Query
:SYSTem:ERRor?
Response
(17 to 25 bytes
+ CRLF)
Error number, "Error description"
Description
Returns and clears the error, if any. When an error occurs, “1” is set for the ERR bit of the
Status Byte Register (STB) until the error is read by this command or
*CLS
is executed.
An instrument error, however, is not cleared. When an instrument error occurs, the
EXT. I/O ERROR terminal output is turned ON. For details about the error, see “Before
requesting repair” (p. 195).
Example
:SYSTem:ERRor?
-388,"Adjust
␣
data
␣
lost"
The adjustment data of the instrument is corrupted.
Measurement control
Starting the test
Syntax
Command
:STARt
Description
Starts the test.
An execution error occurs if this command is executed when the test cannot be started.
For instance, an execution error occurs while a test is being performed, an interlock is
activated, the EXT. I/O STOP signal is ON, etc.
Example
:STARt
Starts the test.
Ending the test
Syntax
Command
:STOP
Description
Ends the test.
Example
:STOP
Ends the test.
Communications Function
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...