204
Influence of Capacitive Load
Repeatability
Measurement conditions
Connect the resistance load and capacitive load in parallel for measurement. (See the following
figure.)
(Use a film capacitor as a capacitor.)
Instrument
In the following test data, a capacitive load is connected and measurement is performed 100 times
in the 200 M
Ω
range for each voltage value. The variation is calculated based on the average of
the 100 measurements. The measured values are acquired with the capacitor charged.
(The test time is 100 seconds at most.)
Sampling time: 1 PLC
Voltage
Capacitance
25 V
50 V
100 V
250 V
500 V
0.001 µF
±
0.134%
±
0.067%
±
0.202%
±
0.134%
±
0.010%
0.01 µF
±
0.202%
±
0.135%
±
0.270%
±
0.135%
±
0.067%
0.1 µF
±
0.741%
±
0.337%
±
0.404%
±
0.202%
±
0.253%
IMPORTANT
The variation may be different depending on the capacitor type.
Cables used
LOW side
L2132 Unterminated Lead (red)
HIGH side
L2133 Unterminated Lead (black)
Instrument settings
Test voltage
100 V
Resistance range
2000 M
Ω
range
Load
1000 M
Ω
Cable length
Measurement speed
1 m
2 m
3 m
4 m
5 m
1 PLC
±
0.2%
±
0.1%
±
0.3%
±
0.3%
±
0.3%
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...