206
Influence of Noise
Synchronizing the sampling time with the commercial power supply frequency
The sampling time of this instrument is synchronized with the power supply frequency.
When the main frequency setting of the supplied power is set to AUTO, 50 Hz or 60 Hz is
automatically recognized and set.
The main frequency can manually be set to 50 Hz or 60 Hz, however, be aware that the measured
value becomes unstable if the setting is wrong.
Measurement signal with power supply noise superimposed
Ideal measurement
signal (DC)
Integral period
Noise caused by a commercial power supply
Separating the power supply line
It is recommended to prepare a separate power supply for power systems, welding machines, etc.
See “Fig. Conductive noise countermeasure”.
Insert a common mode filter (EMI choke) into the noise transmission path.
Select a common mode filter with high impedance, if possible. Multiple filters can increase the
effect.
Instrument
Controller
(PC, PLC)
Motor
Welding
machine
Inverter
Isolation
Common mode filter
to the transmission path
Power supply line
L
N
PE
Power supply line of
other systems
L
N
PE
Isolation
Separate the power
supply line.
RS-232C
LAN
USB
EXT. I/O
Conductive noise countermeasure
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...