5.7 Patient Leakage Current
Measurement Mode
────────────────────────────────────────────────────
8
Condition of the equipment to
be measured for the current
value
Condition of the equipment to be measured
Single fault condition (110% voltage applied):
9
Allowable value
Current setting of the allowable value (Single fault condition only)
10
Setting display for the
equipment to be measured
The following parameters are displayed:
Settings of the equipment to be measured
Model name, Number, Current consumption monitor value, Line voltage
monitor value
Key
Description of the function
11
(Clear the maximum
value)
Use as necessary to clear unnecessary data and start new measurements.
As long as this key is not pressed, the maximum value is not cleared even after
changing measurement settings in the same measuring mode.
The maximum value is cleared after changing the measuring mode.
12
Opens the allowable value set-up window. (Single fault condition only)
Range of the allowable value: 5
μ
A to 20 mA
For details, refer to Section 5.10.1, "Allowable Value Setting."
13
Opens the polarity and equipment condition set-up window.
1. Select either the
or
mode.
For automatic measurement, refer to Section 5.10.3, "Automatic
Measurement Setting."
2. When in the manual mode, in addition to making measurements while
changing the polarity and condition of the equipment to be measured, the
maximum value data can be printed out or saved.
Power source polarity selection:
NORMAL:
, REVERSE:
14
Opens the measurement network filter set-up window.
Select either filter ON or OFF.
: Network with frequency characteristics
: Network with uninducted resistance of only 1 k
Ω
15
Zero adjustment function
Press this key to perform zero adjustment before measuring.
When zero adjustment is effective,
is displayed on the screen.
For details, refer to Section 5.7.3, "Zero Adjustment (ME Equipment - Patient
Leakage Current
)."
16
Opens the print-out window.
Prints out the maximum value displayed on the screen.
For details, refer to Section 5.10.5, "Printing out the Maximum Values."
17
Opens the save data window.
For details, refer to Section 5.10.6, "Saving the Maximum Values."
18
Exits the patient leakage current
measurement, and returns to the initial
screen.
(2) Settings
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......