![Hioki 3155-01 Instruction Manual Download Page 123](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164123.webp)
103
────────────────────────────────────────────────────
5.10 Condition Set-up Window Setting (ME Equipment)
────────────────────────────────────────────────────
5.10.7 Low Resistance Measurement with 3157
NOTE
Low resistance measurement mode allows measurement with optional 3157
AC GROUNDING HiTESTER (maximum 31 AAC, in accordance with multiple
standards). Using the RS-232C interface enable the 3155 to control the 3157.
Prior to set up, connect 3157 to equipment to be measured and 3157 to 3155
with connection cable. Synchronize 3157 communication settings with 3155.
3157 communication settings are applied when power is turned ON. Set up
and configure while power is OFF.
See "When using 3157" on the previous page.
See also Section 8.5.1, "Communication Conditions Setting."
Settings
(1) On the low resistance measurement mode
screen, press the
key to open the
measurement window.
(2) Press the key for the desired measuring
equipment.
1.
: Select 3155.
2.
: Select 3157.
When 3157 is selected, the
key is
displayed. Measurement and data transfer
status are displayed.
(3) Set up 3157 test settings. Measurement does not start unless the following
conditions are met.
1. Test settings
・
Unit of the maximum and minimum test values: Resistance
・
Test time: ON
・
Maximum test value: ON
When the optional minimum test value setting function is ON.
・
Minimum test value: OFF
2. Optional function setting
Endless timer function: Not set
(4) Verify 3157 is set to READY. Press the
key to start measurement.
Automatically 3157 maximum test value is loaded as an allowable value.
After measuring, measurement results are displayed. Unless 3157
judgment is either PASS or FAIL (UPPER FAIL), test results are not
displayed.
(5) Press the
key to close the measurement window.
・
For usage, connection and communication settings for the 3157, see
3157/3157-01 AC GROUNDING HiTESTER Instruction Manual and 9593-02
RS-232C INTERFACE Instruction Manual.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......