![Hioki 3155-01 Instruction Manual Download Page 126](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164126.webp)
106
────────────────────────────────────────────────────
5.11 System Screen (ME Equipment)
────────────────────────────────────────────────────
5.11.2 Saved Data Reference Screen (ME Equipment)
In the example shown in the figure, there
are a total of three data units, of which
the one being displayed is the second.
The data shown is for enclosure leakage
current measurement.
NOTE
The following is an explanation of how to use the saved data.
Up to 100 data units can be saved. However, up to 2000 data can be saved.
For the model name and number settings, refer to Section 5.2.2, "Equipment
Set-up Screen (ME Equipment)."
(1) Settings
1. On the initial screen, press the
key to move to the system
screen.
2. On the system screen, press the
key to move to the saved
data reference screen.
(2) Saved data reference screen display
On the saved data reference screen, saved data can be viewed, deleted or
printed out.
Number of current data unit / total number of data
units
Model name
Number
Measuring modes
(up to 7 entries)
Earth Leak
: Earth leakage current
measurement
Encl. Leak:
Enclosure leakage current
measurement
Resistance
: Low resistance measurement
Pat.Leak
:
Patient leakage current
measurement
Pat.Leak
:
Patient leakage current
measurement
Pat.Leak
:
Patient leakage current
measurement
Pat.Aux:
Patient auxiliary current
measurement
Maximum value
Maximum value judgment
Condition at time of maximum value judgment
Polarity
Condition of the equipment to be measured
Measurement current setting
Measurement network filter setting
Contact condition setting
Measuring equipment in low resistance
measurement mode
When the maximum value is 1
μ
A or lower, "UNDER 1
μ
A" is displayed.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......