265
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Queries the maximum values after automatic measurement.
Syntax
:MEASure:AUTO?_<
data>
Function
Returns the measurement results collectively.
Returns the maximum values of the combinations of the power source
polarity and condition of the equipment to be measured, their judgments,
power source polarities, and conditions of the equipment to be measured, as
numerical values in NR3 and NR1 format.
The data are ordered as follows: <
maximum value>, <
judgment>, <
power source polarity> and <
condition of the equipment to be
measured>, and the available combination of the automatic measurement is
read in a format of
to
.
<Judgment>
(If the allowable value judgment is disabled, 0)
Allowable value or lower (PASS): 0
Higher than allowable value (FAIL): 1
<Power source polarity>
(If the equipment to be measured is set to the internally powered
equipment, 0)
Normal polarity: 0
Reverse polarity: 1
<Condition of the equipment to be measured>
Normal condition: 0
Single fault condition (open power lead): 1
Single fault condition (open ground): 2
Single fault condition (110% voltage applied): 3
Note
Execute this command after verifying that the automatic measurement is
finished (:AMC? command). Executing this command during automatic
measurement will not obtain the correct result.
Use this command only after using the ":MODE" command to set the
measuring mode.
:MEASure:AUTO?
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......