249
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Sets the allowable value.
Syntax
When the measuring mode has one allowable value:
:CONFigure:COMParator_<
data>
When the measuring mode has two allowable values:
:CONFigure:COMParator_<
data1>,
<
data2>
data
<data> Allowable value (Numerical value in NR3 format)
<data1> Allowable value in normal condition (Numerical value in NR3
format)
<data2> Allowable value in single fault condition (Numerical value in NR3
format)
Function
Sets the allowable value.
The main unit displays allowable value as a product of the numerical value
and the coefficient, but when using RS-232C communications, the
coefficient is set as 100%. When setting the low resistance measurement,
the unit is
Ω
and the range is 10.00E-03 to 500.0E+00. Otherwise the unit
is A and the range is 5.000E-06 to 20.00E-03.
The patient leakage current
and
, and low resistance measurement
modes have one allowable value.
Sets one allowable value.
The earth leakage current, enclosure leakage current, patient leakage
current
, and patient auxiliary current measurement modes have two
allowable values.
Sets each allowable value in normal condition and single fault condition.
Note
Use this command only after using the ":SYSTem:COMParator" command
to enable allowable value judgment, and the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:COMParator 0.5E‑3
The allowable value is set to 500.0
μ
A.
Transmission
:CONFigure:COMParator 0.1E‑3,0.5E‑3
The allowable value in normal condition is set to 100.0
μ
A, and the
allowable value in single fault condition is set to 500.0
μ
A.
Error
(ME equipment) (ordinary equipment)
If the allowable value judgment is disabled, or if the measuring mode has
not been set, or if the numerical value exceeding the range has been set, an
error occurs.
If the number of allowable values that can be set is not equal to the
number of data, an error occurs.
:CONFigure:COMParator
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......