10.2 Fuse Replacement
────────────────────────────────────────────────────
10.2.3 Replacement of the 9196 APPLY UNIT Power Source
Fuse
WARNING
To prevent electric shock when replacing the power source fuse,
always turn the power switch off and disconnect the power cord and
probe before beginning.
Only use fuses of the specified type that is rated for the specified
current and voltage. Using a fuse that does not meet the specifications
or shorting the fuse holder may cause an accident that might result in
injury or death. Specified fuse: 250 V T0.25 AL 20 mm
×
5 mm dia.
Opening and closing the fuse holder
Opening
: Rotate it counter-clockwis
e.
Closing
: Rotate it clockwise.
Fuse holder
The fuse holder contains
the fuse.
The power source fuse is on the back of the 9196.
Replace the blown fuse as follows. (See Figure.)
1. Turn the power OFF, and disconnect the power cord and probe.
2. Using a Phillips screwdriver, rotate the fuse holder counter-clockwise to
open.
3. Replace the fuse with an equivalent replacement.
4. Replace the fuse holder, and rotate it clockwise to close.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......