102
────────────────────────────────────────────────────
5.10 Condition Set-up Window Setting (ME Equipment)
────────────────────────────────────────────────────
・
A part of the condition of the equipment to be measured is omitted
depending on the settings of the grounding class and applied part type of
the equipment to be measured.
・
In the following cases, model name and number must be changed for storage
as a different data unit.
・
When the measurement network filter setting is changed.
・
When the measurement current setting is changed.
・
When any of the following conditions pertain enclosure leakage current.
When the contact condition setting is changed.
When 110% voltage (normal polarity) is applied.
When 110% voltage (reverse polarity) is applied.
・
When the polarity of applied 110% voltage is changed in patient
leakage
current
.
・
When the polarity of applied 110% voltage is changed in patient
leakage
current
.
Notes when saving data
・
Even if the filter setting, measurement current and contact condition have
changed, as long as the registered data unit is the same, the data for that
data unit will be overwritten. (after a request for confirmation)
・
When saving data manually for each setting condition (power source polarity
and condition of the equipment to be measured), clear the maximum value
prior to each measurement and save after changing the model name and
number. For automatic measurement, refer to Section 5.10.3, Automatic
Measurement Setting (ME Equipment)."
・
For changing the model name and number, refer to Section 5.2.2,
"Equipment Set-up Screen."
Settings
(1) On the measuring mode screen, press the
key.
(2) At the bottom of the screen, the save data window will
open.
Press the
key to save the maximum values.
If a data unit with the same name already exists, a
request for confirmation appears.
Press the
key to close the save data window
without saving.
If the
key is pressed on the measuring mode
screen without first either saving or printing out the
current data, that data will be lost.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......