282
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Queries errors.
Syntax
:SYSTem:ERRor?
Function
Returns the value of error register as a numerical value in NR1 format, and
then clears error register.
Note
For the error register, refer to Section 7.5.11, "Error Register."
Example
Transmission
:SYSTem:ERRor?
Response
Headers ON
:SYSTem:ERRor 64
Headers OFF
64
A command error has occurred.
■
Sets the language displayed on the screen.
Syntax
:SYSTem:LANGuage_<
data>
data
JAPanese/ENGlish
Function
Sets the language displayed on the screen.
JAPanese Sets the display language to Japanese.
ENGlish Sets display language to English.
Example
Transmission
:SYSTem:LANGuage JAPanese
The display language is set to Japanese.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has been selected, an error occurs.
:SYSTem:ERRor?
:SYSTem:LANGuage
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......