5.5 Patient Leakage Current
Measurement Mode
────────────────────────────────────────────────────
5.5.2 Connecting Equipment for Patient Leakage Current
Measurement
CAUTION
・
The maximum current that can be input into the leakage current
measurement terminals of the 3155 is 25 mA (AC+DC). The input of any
larger current will cause the beeper to sound (when the beeper is turned on),
after which continued input may cause the fuse in the measurement network
to fail. In this case, input must be terminated immediately.
Afterwards, a fuse check is performed automatically upon returning to the
initial screen from the leakage current measurement mode. Therefore, do
not connect anything to leakage current measurement terminal T1. (If
anything is connected, the display may show "Measurement Network Fuse
blown.")
・
The rated capacity of the auxiliary power socket is 1500 VA or less. If this
capacity is exceeded, the auxiliary power outlet circuit breaker will actuate,
and the power supply to the equipment to be measured will be cut off.
(With the 3155-01, the main power switch is turned off.)
・
When power of the equipment to be measured exceeds 1500VA, make the
connection directly to the outlet; do not use the auxiliary power socket.
(When measuring class
equipment, connect the protective earth conductor
of the equipment to be measured to the G terminal of the 3155.)
In this case, the power source polarity and the condition of the equipment to
be measured cannot be set on the 3155.
・
Use an insulating transformer.
Applied part
Red 9170 TSET LEAD
Equipment to be measured
Insulating transformer
(1) Class
and class
equipment
1. Plug the power cord of the equipment to be measured into the auxiliary
power socket found on the right side of the 3155.
2. Connect the red 9170 TEST LEAD earth to the leakage current
measurement terminal T2.
3. Touch the red 9170 TEST LEAD to the applied part of the equipment to be
measured.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......