132
────────────────────────────────────────────────────
6.3 Enclosure Leakage Current Measurement Mode (for IEC/TR 60990)
────────────────────────────────────────────────────
6.3.4 Measurement Accompanying IEC 61010-1 (1990-09) + am1
(1992-09) + am2 (1995-07)
NOTE
Item
IEC 61010-1
Measurement method
Manual /automatic
Power source polarity
Normal polarity / reverse polarity
Contact condition
Between the enclosure and the earth
Condition of the equipment
to be measured
Filter setting Measurement
current
Allowable
value
Measuring object
Normal condition
ON1
AC+DC
0.5 mA
Sinusoidal waveforms
AC peak
0.7 mA
Non-sinusoidal waveforms
or mixed frequencies
DC
2 mA
Single fault
condition
Open power
lead
ON1
AC+DC
3.5 mA
S
inusoidal waveforms
AC peak
5 mA
Non-sinusoidal waveforms
or mixed frequencies
DC
15 mA
Open ground
ON1
AC+DC
3.5 mA
Sinusoidal waveforms
AC peak
5 mA
Non-sinusoidal waveforms
or mixed frequencies
DC
15 mA
NOTE
IEC/TR 60990 (1990-06) sets standard measurement methods for current. IEC
60950 (1991-10) + am4 (1996-07), IEC 61010-1 (1990-09) + am1 (1992-09) +
am2 (1995-07), etc. use the methods of IEC/TR 60990 (1990-06).
When performing current measurements with the 3155 intended to conform to
IEC 61010-1 (1990-09) + am1 (1992-09) + am2 (1995-07), use the enclosure
leakage current measurement mode. Plug the power cord of the equipment to
be measured into the auxiliary power socket, and touch the test point with the
9170 TEST LEAD.
Test points taken between the accessible part and the reference test earth
are defined as between the enclosure and the earth.
・
In the enclosure leakage current measurement mode, the 50
μ
A range is
not used.
・
When testing class 0I or other older models, use (UL1015, AWG #18 or
equivalent) wire to connect the G terminal with the protective earth
terminal of the equipment to be measured.
Measurement screen settings
Measuring mode: Enclosure leakage current
Since the maximum current that can be input into the leakage current
measurement terminals of the 3155 is 25 mA (AC+DC), if the measurement
network filter setting is OFF, measurements are impossible up to the standard
allowable values.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......