254
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Sets the measurement current.
Syntax
:CONFigure:CURRent_<
data>
data
ACDC/AC/DC/ACPeak
Function
Sets the measurement current.
(ME equipment)
9497 NETWORK B
ACDC
Sets the measurement current to AC+DC.
AC
DC
(Ordinary equipment)
9498 NETWORK C
ACDC
Sets the measurement current to AC+DC.
AC
DC
ACPeak AC peak (The measurement network filter is automtically set to
ON1.)
9499 NETWORK D
ACDC
Sets the measurement current to AC+DC.
AC
DC
ACPeak AC peak
Note
For ME equipment, use this command only after using the ":MODE"
command to set the measuring mode to the patient leakage current
or
patient auxiliary current.
For ordinary equipment, use this command only after using the ":MODE"
command to set the measuring mode.
Example
Transmission
:CONFigure:CURRent ACDC
The measurement current is set to AC+DC.
Error
(ME equipment)
If the measuring mode has not been set, or when in a mode other than the
patient leakage current
mode or patient auxiliary current mode, an error
occurs.
(Ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, an error occurs.
:CONFigure:CURRent
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......