3.2 9196 APPLY UNIT
────────────────────────────────────────────────────
3.2.1 9196 APPLY UNIT
NOTE
3.2 9196 APPLY UNIT
The 9196 APPLY UNIT is an instrument which outputs 110% of the input
power voltage.
Also, it has a function to break the circuit to prevent damage to the equipment
to be measured and electric shock to the user if an output current exceeding
6 mA.
Combined with the 3155 LEAK CURRENT HiTESTER, it is used to test
leakage current in medical electrical equipment. Leakage current from
medical electrical equipment has the greatest potential for harming human
beings, and is therefore strictly regulated even with regard to single fault
conditions. The following 6 types of testing for leakage current in medical
electrical equipment are available.
Earth leakage current
Enclosure leakage current
Patient leakage current
Patient leakage current
Patient leakage current
Patient auxiliary current
The 9196 is used in testing for types
,
, and
.
For the fuse replacement, refer to Section 10.2.3, "Replacement of the 9196
APPLY UNIT Power Source Fuse."
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......