6.6 Condition Set-up Window Setting (for IEC/TR 60990)
────────────────────────────────────────────────────
6.6.4 Measurement Current Setting (for IEC/TR 60990)
NOTE
This is where the measurement current is set.
Settings
(1) Press the
key on the leakage current measurement
mode screen to open the measurement current set-up
window.
Measurement continues even when this window is open.
(2) On the measurement current set-up window, press the key
for the type of current you wish to use.
: AC rms
: DC
: AC+DC rms
: AC peak
(3) Press the
key to close the measurement current
set-up window.
・
When set to measurement current AC peak, the measurement network filter
cannot be set to OFF. If set to OFF, it automatically changes to ON1.
・
When set to measurement current DC, the measurement network filter cannot
be set to ON2. If set to ON2, it automatically changes to OFF.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......