![Hioki 3155-01 Instruction Manual Download Page 100](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164100.webp)
80
────────────────────────────────────────────────────
5.8 Patient Auxiliary Current Measurement Mode
────────────────────────────────────────────────────
Black 9170 TEST LEAD
Equipment to
be measured
Insulating transformer
Red 9170 TEST LEAD
Applied parts
NOTE
・
When the auxiliary power socket does not fit the power plug of the
equipment to be measured, use the supplied power plug.
・
When testing class 0I or other older models, use (UL1015, AWG #18 or
equivalent) wire to connect the G terminal with the protective earth
terminal of the equipment to be measured.
・
Use wire for connecting the functional earth terminals of an equipment to
be measured to the ground.