5.3 Earth Leakage Current Measurement Mode (ME Equipment)
────────────────────────────────────────────────────
5.3.1 Earth Leakage Current Measurement Mode
Condition of the equipment to be measured
Power source polarity
Normal condition
Normal polarity
Reverse polarity
Single fault condition
Open power lead
Normal polarity
Reverse polarity
NOTE
Class
equipment
Type B applied part
Type BF applied part
Type CF applied part
Normal
condi-
tion
Single fault condition Normal
condi-
tion
Single fault condition
Normal
condi-
tion
Single fault condition
Open
power
lead *1
Open
power
lead
Open
power
lead
Earth
leakage
current
General
equipment
0.5mA
1mA
−
−
0.5mA
1mA
−
−
0.5mA
1mA
−
−
Special
equipment
and mobile
equipment
2.5mA
5mA
−
−
2.5mA
5mA
−
−
2.5mA
5mA
−
−
Permanently
installed
equipment
5mA
10mA
−
−
5mA
10mA
−
−
5mA
10mA
−
−
5.3 Earth Leakage Current Measurement Mode (ME
Equipment)
Earth leakage current measurement for medical electrical equipment is
performed only when class
equipment has been set. When either class
or
internally powered equipment has been set, the earth leakage current key on
the initial screen is non-selectable.
Measurement is possible only when the power cord of the equipment to be
measured has been plugged into the auxiliary power socket of the 3155.
The following combinations are measured under the earth leakage current
measurement.
The measurement current is set to "AC+DC rms."
For the concept of leakage current in medical electrical equipment and its
wiring when using the 3155, refer to Appendix 2.1, "IEC 60601-1 (1988-12) +
am1 (1991-11) + am2 (1995-03)."
By using the automatic measurement mode, it is possible to measure changing
automatically the power source polarity and condition of the equipment to be
measured, and obtain the maximum value of the combination. For details,
refer to Section 5.10.3, "Automatic Measurement Setting."
Measurement types and their allowable values
*1: Interruption of one supply conductor
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......