91
────────────────────────────────────────────────────
5.10 Condition Set-up Window Setting (ME Equipment)
────────────────────────────────────────────────────
5.10.1 Allowable Value Setting (For Leakage Current
Measurement)
Screen during input of the value in normal
condition
(Numerals are displayed in the input cell.)
Set value = numeral
×
coefficient
Range: 5
μ
A to 20 mA
NOTE
5.10 Condition Set-up Window Setting (ME Equipment)
The allowable value setting may be revised as necessary to accommodate the
measuring mode and condition of the equipment to be measured (normal
condition or single fault condition).
The 3155 can set two allowable values in normal condition and single fault
condition.
When shipped, the allowable values are set in accordance with IEC 60601-1
(1988-12) + am1 (1991-11) + am2 (1995-03).
For details, refer to Section 10.4.2, "Default Settings."
Settings
(1) On the measuring mode screen, press the
key to open the
allowable value set-up window.
Current setting
Input numeral
Input coefficient
Keys for numerical input
Normal: Press to set the allowable value in normal condition
Fault: Press to set the allowable value in single fault condition
Numerals: Press to input numerals
Coefficient: Press to input coefficient
Units: Highlighted in white during numerical input. Press after numerical
input to finalize units.
Return: Press to close the allowable value set-up window.
If a value of less than 5
μ
A is input, the value is automatically revised to
5
μ
A. If a value of more than 20 mA is input, the value is automatically
revised to 20 mA.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......