74
────────────────────────────────────────────────────
5.7 Patient Leakage Current
Measurement Mode
────────────────────────────────────────────────────
5.7.4 Basic Settings for Patient Leakage Current
Measurement
Patient leakage current
measurement
mode screen
Polarity and equipment condition
set-up window
Initial screen
Equipment set-up screen
Select and press the key for the power
source polarity to set.
Measurement continues even while the
window is open.
For automatic measurement, refer to
Section 5.10.3, "Automatic Measurement
Setting."
Measurement starts automatically
Allowable value set-up window
Measurement network filter set-up
window
Set the filter.
Set the allowable value referring to Section
5.10.1, "Allowable Value Setting."
(For enabling or disabling the allowable value
judgment, refer to Section 5.11.4, "Allowable
Value Judgment Set-up Screen.")
Set the grounding class of the
equipment to be measured.
Select the type BF or CF applied
part.
This illustrates basic settings and reference categories until measurement.
Preparations for measurement follow Section 5.7.2, "Connecting Equipment for
Patient Leakage Current
."
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......