10.4 System Reset
────────────────────────────────────────────────────
NOTE
Settings for patient leakage current
Measurement method Manual
Condition of the equipment
to be measured
Single fault condition
(110% voltage applied)
Power source polarity Normal
Measurement network filter
ON (frequency
characteristics)
Class
equipment
Type B applied part
Type BF applied part
Type CF applied part
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Earth leakage current
0.5 mA
1 mA
0.5 mA
1 mA
0.5 mA
1 mA
Enclosure leakage current
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.1 mA
0.5 mA
Patient leakage
current
DC
0.01 mA
0.05 mA
0.01 mA
0.05 mA
0.01 mA
0.05 mA
AC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
AC+DC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
Patient leakage current
5 mA
Patient leakage current
5 mA
0.05 mA
Patient auxiliary
current
DC
0.01 mA
0.05 mA
0.01 mA
0.05 mA
0.01 mA
0.05 mA
AC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
Class
equipment
Type B applied part
Type BF applied part
Type CF applied part
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Enclosure leakage current
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.1 mA
0.5 mA
Patient leakage
current
DC
0.01 mA
0.05 mA
0.01 mA
0.05 mA
0.01 mA
0.05 mA
AC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
AC+DC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
Patient leakage current
5 mA
Patient leakage current
5 mA
0.05 mA
Patient auxiliary
current
DC
0.01 mA
0.05 mA
0.01 mA
0.05 mA
0.01 mA
0.05 mA
AC
0.1 mA
0.5 mA
0.1 mA
0.5 mA
0.01 mA
0.05 mA
Internally powered
equipment
Type B applied part
Type BF applied part
Type CF applied part
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Normal
condition
Single fault
condition
Enclosure leakage current
0.1 mA
0.1 mA
0.1 mA
Patient leakage
current
DC
0.01 mA
0.01 mA
0.01 mA
AC
0.1 mA
0.1 mA
0.01 mA
AC+DC
0.1 mA
0.1 mA
0.01 mA
Patient leakage current
5 mA
Patient leakage current
5 mA
0.05 mA
Patient auxiliary
current
DC
0.01 mA
0.01 mA
0.01 mA
AC
0.1 mA
0.1 mA
0.01 mA
・
Single fault condition (110% voltage applied) is performed with the optional
9196 APPLY UNIT etc.
・
After a system reset, because the default applied part type setting is type B,
the patient leakage current
measurement mode is inoperable. After
changing the applied part type setting, the following default settings are
effected:
Allowable values when the 9497 NETWORK B is installed
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......