3.2 9196 APPLY UNIT
────────────────────────────────────────────────────
Rear view
Voltage application button
9190 VOLTAGE APPLY PROBE
Voltage application probe connection terminal (OUTPUT)
Connect the 9190 VOLTAGE APPLY PROBE.
Output monitor (OUTPUT MONI): Red monitor lamp
The condition of the output to the voltage application probe is shown by this
monitor lamp.
If illuminated, the output is ON. (The voltage application button to the voltage
application probe is being held down.)
If not illuminated, the output is OFF.
Handle
Fuse
Use the HIOKI specified fuse. (250 V T0.25 AL 20 mm
×
5 mm dia.)
Power inlet
The rated power voltage of the 9196 is 100 to 240 VAC (45 to 400 Hz), and the
maximum rated power is 25 VA.
Connect a grounded three-core power cord that is supplied with the unit.
Function earth terminal
If the protective ground terminal cannot be connected to the ground via a
grounded three-core power cord supplied, connect the function earth terminal
to the ground.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......