117
────────────────────────────────────────────────────
6.1 Measurement Network C
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
6.1.1 Leakage Current Measurement Types and Their Allowable
Values for Ordinary Electrical Equipment
NOTE
The 3155 unit contains an allowable value judgment function, which compares
the maximum value to the allowable value.
To use this function, set the judgment function to ON, and set each of the
allowable values.
(1) Leakage current measurements for ordinary electrical equipment
The 3155 can perform the following leakage current measurements:
・
Earth leakage current (for class
equipment only)
・
Enclosure leakage current
(2) Allowable values
The allowable values vary per the applicable regulations.
Example:
Allowable values per IEC 60950 (1991-10) + am4 (1996-07)
・
Class
equipment (hand-held): 0.75 mA
・
Class
equipment (movable (other than hand-held)): 3.5 mA
・
Class
equipment (stationary, pluggable type A): 3.5 mA
・
Class
equipment (stationary, permanently connected or pluggable type B)
Equipment with earth leakage current not exceeding 3.5 mA: 3.5 mA
Equipment with earth leakage current exceeding 3.5 mA: 5% of input
current
・
Class
equipment: 0.25 mA
Allowable values vary per the applicable laws and regulations, but when using
the 9498 NETWORK C (conforms to IEC/TR 60990), the default value is 3.5
mA (0.25 mA for other than class
equipment). Set the allowable value as
necessary.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......