296
────────────────────────────────────────────────────
9.1 General Specifications
────────────────────────────────────────────────────
Dimensions and Mass
212
W
×
292H
×
110D mm(8.35"W
×
11.5"H
×
4.33"D) (excluding
projections), Approx. 4.5 kg (15.9 oz.)
Accessories
Grounded three-core power cord
9170 TEST LEADS
Instruction Manual
9195 ENCLOSURE PROBE
Spare fuse (3155 120 V: 250 V T0.5 AL)
(3155 200/240 V: 250 V T0.25 AL)
(3155-01: 250 V T0.2 AL)
9233 RECORDING PAPER (1 roll)
9399 CARRYING CASE (for accessories)
Alligator clips (red and black)
Outlet power plug (3155 200/240 V and 3155-01 only)
Voltage selector key (3155-01 only) 2
Options
9497 NETWORK B (for medical electrical equipment)
9498 NETWORK C (for IEC/TR 60990)
9499 NETWORK D (universal)
9461 PIN-TYPE LEADS
9287 CLIP-TYPE LEADS (do not conform to IEC 61010-2-031)
9233 RECORDING PAPER (10 m, 10 rolls)
9196 APPLY UNIT (for patient leakage current
and
measurements,
and part of enclosure leakage current measurement)
9190 VOLTAGE APPLY PROBE (attached to the 9196)
9388 CARRYING CASE (with casters)
Standards applying
EMC EN55011:1991
EN50082-1:1992 *
EN61000-3-2:1995
EN61000-3-3:1995
Safety EN61010-1:1993+A2:1995
Pollution Degree 2, Overvoltage Category
(anticipated transient overvoltage 2500 V)
EN61010-2-031:1994
Recording method
Thermal serial printer
Number of dots
16 dots
Recording paper
58 mm
×
10 m
Effective recording width
47 mm
Maximum printing speed
41 characters/s
Printed data
1) Data on the equipment to be measured (model name, number and
grounding class), 2) Measuring mode, 3) Maximum value, 4) Settings, 5)
Judgment
Data stored
Data on the equipment to be measured
Storage capacity
100 units
Interface
RS-232C (1ch)
Connector
D-subminiature 9-pin plug (male)
*Note: Accurate measurement may be impossible in locations subject to strong external
electromagnetic fields.
Measurement data by HIOKI:
Electromagnetic field intensity 3V/m (at 27 M to 500 MHz)
Leakage current measurement ac20% rdg. (25 mA range,10 mADC input)
Low resistance measurement ac30% rdg. (50
Ω
range, 10
Ω
input)
[Recording]
[Storage]
[Interface]
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......