201
────────────────────────────────────────────────────
7.5 Low Resistance Measurement Mode (Universal)
────────────────────────────────────────────────────
7.5.2 Connecting Equipment for Low Resistance Measurement
(Universal)
WARNING
Do not input a voltage exceeding 30 Vrms, 42.4 Vpeak or 50 VDC between
the resistance measurement terminals, and between the resistance
measurement terminal and the ground.
CAUTION
・
Be careful not to hurt yourself on the end of the 9461 PIN-TYPE LEAD,
which is pointed.
・
Do not connect anything other than the 9461 PIN-TYPE LEAD and the 9287
CLIP-TYPE LEAD to the resistance measurement terminal.
NOTE
Black 9461 PIN-TYPE LEAD
Equipment to
be measured
Red 9461 PIN-TYPE LEAD
Do not make measurements if the power cord of the equipment to be
measured is plugged into the auxiliary power outlet of the 3155.
(1) When using 9461 PIN-TYPE LEADS
The 9461 PIN-TYPE LEADS are connected as shown in the figure. Connect to
all four terminals.
Do not connect anything to the auxiliary power outlet of the 3155.
Touch the pin-type leads between the protective earth terminal or earth
contact and the part required to be earthed.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......