![Hioki 3155-01 Instruction Manual Download Page 154](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164154.webp)
134
────────────────────────────────────────────────────
6.3 Enclosure Leakage Current Measurement Mode (for IEC/TR 60990)
────────────────────────────────────────────────────
Red 9170 TEST LEAD
Equipment to
be measured
9195
NOTE
When the auxiliary power socket does not fit the power plug of the equipment
to be measured, use the supplied power plug.
If the enclosure or any part thereof is made of material with insulating
characteristics
If the enclosure or any part thereof is made of material with insulating
characteristics, attach the 9195 ENCLOSURE PROBE to the enclosure at
approximately 0.5 N/cm
2
across the entire surface.