![Hioki 3155-01 Instruction Manual Download Page 279](http://html1.mh-extra.com/html/hioki/3155-01/3155-01_instruction-manual_2137164279.webp)
259
────────────────────────────────────────────────────
8.8.2
Commands
────────────────────────────────────────────────────
■
Sets the power source polarity in manual measurement.
Syntax
:CONFigure:POLarity_<
data>
data
NORMal/REVerse
Function
Sets the power source polarity in manual measurement.
NORMal Sets to the normal polarity.
REVerse Sets to the reverse polarity.
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:AUTO OFF;:CONFigure:POLarity NORMal
The power source polarity is set to the normal polarity.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, or when in automatic measurement mode, or if the
equipment to be meaured is set to the internally powered equipment, an
error occurs.
■
Queries the power source polarity in manual measurement.
Syntax
:CONFigure:POLarity?
Function
Queries the power source polarity setting in manual measurement.
<data>
NORMAL The power source polarity has been set to the normal polarity.
REVERSE
The power source polarity has been set to the reverse polarity.
Note
Use this command only after using the ":MODE" command to set the
measuring mode.
Example
Transmission
:CONFigure:AUTO OFF;:CONFigure:POLarity?
Response
Headers ON
:CONFIGURE:POLARITY NORMAL
Headers OFF
NORMAL
The power source polarity has been set to the normal polarity.
Error
(ME equipment) (ordinary equipment)
If the measuring mode has not been set, or when in the low resistance
measurement mode, or when in automatic measurement mode, or if the
equipment to be meaured is set to the internally powered equipment, an
error occurs.
:CONFigure:POLarity
:CONFigure:POLarity?
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......