5.10 Condition Set-up Window Setting (ME Equipment)
────────────────────────────────────────────────────
5.10.3 Automatic Measurement Setting (ME Equipment)
NOTE
Range Standby time: 0 to 99 seconds
Measuring time: 1 to 99 seconds
NOTE
In each of the leakage current measurement modes, automatic measurement
can be performed, in which the polarity and condition of the equipment to be
measured are automatically changed. The combinations shown below are all
set, and the maximum value of each combination is measured.
After measurement, the results will be displayed on the automatic
measurement result screen. Printing and saving are possible on this screen.
(1) Power source polarity (Normal and reverse)
(2) Condition of the equipment to be measured
Normal condition
Single fault condition (Open power lead, open ground)
Automatic measurement is not possible for 110% voltage application. Use the
optional 9196 APPLY UNIT to make this measurement manually.
Settings
(1) On the leakage current measurement mode
screen, press the
key to open the polarity
and equipment condition set-up window.
Measurement continues even when this window
is open.
(2) Press the key for the desired measurement
method.
1.
: Sets the measurement method to manual.
2.
: Sets the measurement method to
automatic.
After pressing the automatic key, the measuring
time and standby time set-up window opens.
(3) Set the measuring and standby times.
Use the
and
keys to make the
settings as desired.
(4) Press the
key to close the measuring time
and standby time set-up window.
The equipment to be measured may take time to stabilize after changing the
polarity and equipment condition. Set the standby time considering this time.
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......