5.9 Low Resistance Measurement Mode (ME Equipment)
────────────────────────────────────────────────────
SOURCE
SENSE
When not in use
During measurement
NOTE
When the power cord
is removable.
When the power cord
is not removable.
Black 9287 CLIP -TYPE LEAD
Red 9287 CLIP-TYPE LEAD
Equipment to
be measured
As shown in the figure, the 9461 PIN-TYPE
LEAD has a SENSE sheathed by the
SOURCE. When touched to the equipment
to be measured, both the SENSE and the
SOURCE must be in contact with the
equipment to be measured.
(2) When using the 9287 CLIP-TYPE LEADS
The 9287 CLIP-TYPE LEADS are connected as shown in the figure. Connect
to all four terminals.
Do not connect anything to the auxiliary power outlet of the 3155.
1. When the power cord is removable.
Touch the clip-type leads between the protective earth terminal and the
enclosure which is not protectively earthed.
Allowable value: 100 m
Ω
2. When the power cord is not removable.
Touch the clip-type leads between the protective earth pin of the power plug
and the enclosure which is not protectively earthed.
Acceptable value: 200 m
Ω
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......