5.2 Screens (ME Equipment)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
5.2.1 Initial Screen (ME Equipment)
5.2.2 Equipment Set-up Screen (ME Equipment)
NOTE
NOTE
When the power is first turned on, the basic screen for
controlling the 3155 immediately appears, called the
initial screen.
Equipment set-up
: The grounding class and applied part
type of the equipment to be measured
Measuring mode
: Earth leakage current, enclosure
leakage current, patient leakage
current
, patient leakage current
,patient leakage current
,
patientauxiliary current, low resistance
measurement
System
: Clock, beeper, RS-232C communications etc.
On the initial screen, press
the
key to open the
equipment set-up screen.
On the equipment set-up screen, settings for the
grounding class and applied part type of the equipment to
be measured, and registration of the model name and
number can be made.
Press the
key to return to the initial screen.
Grounding class setting
: Class
equipment
: Class
equipment
: Internally powered equipment
If changing the setting for the grounding class of the equipment to be
measured, the condition settings on each measuring mode screen are
initialized, except for the coefficient of the allowable value.
Applied part type setting
: Type B applied part
: Type BF applied part
: Type CF applied part
If changing the setting for the applied part type of the equipment to be
measured, the allowable value on each measuring mode screen is initialized.
For details, refer to Section 5.10.1, "Allowable Value Setting (For Leakage
Current Measurement)," and Section 5.10.2, "Allowable Value Setting (For
Low Resistance Measurement)."
Summary of Contents for 3155-01
Page 1: ...LEAK CURRENT HiTESTER INSTRUCTION MANUAL 3155 01...
Page 2: ......
Page 12: ......
Page 27: ...7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 2 Names and Functions of Parts...
Page 45: ...25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 A Chapter 4 Preparations for Measurement...
Page 243: ...223 7 7 System Screen Universal Example of a data unit printout Example of a details printout...
Page 314: ...294...
Page 322: ...302 9 3 Accuracy Tables...
Page 334: ...314 10 6 External Dimensions 292 2 110 2 212 2 115 8 2 10 6 External Dimensions...
Page 335: ...315 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 APP Appendices...
Page 362: ...INDEX 6 Index...
Page 363: ......
Page 364: ......
Page 365: ......