4-14
NOTE
The vaporizer has deviations due to change of barometric pressure (high altitude).
And the Riken F-211 gas analyzer also has deviations. When testing the vaporizers
using the Riken F-211 gas analyzer, the altitude can be ignored as the deviations
cancel each other out. If using a different gas analyzer, check the effect of change of
barometric pressure before using it in high altitude area.
4.9.4 Vaporizer Leak Test
1. Set the Auto/Manual switch to the Manual position.
2. Set the ACGO to the patient circuit (when configuring the ACGO).
3. Set the APL valve to the SP position.
4. Connect the manual bag to the bag arm port.
5. Connect the Y-piece of the breathing circuit to the leak test port and both ends to the
inspiratory/expiratory port.
Note: When the AG module is configured, the sampling line of the module sample gas should not
be connected to the Y-piece sampling port and ensure that the Y-piece sampling port is blocked.
6. Mount and lock the vaporizer onto the vaporizer mount. (Certain vaporizers need to be
set to at least 1% for correct testing. See the vaporizer manufacturer's manual for
details.)
7. Set the fresh gas flow to 0.2 L/min.
8. Set the APL valve to 75 cmH2O.
9. Press the quick oxygenation button so that the pressure on the PAW gauge increases
above 30 cmH2O
.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...