5-22
5.15 ACGO
ACGO has two options, that is, Off, ACGO with three-way valve. Among them, Off applies to the
machine that does not need to be configured with ACGO, machine that needs to be configured
with a non-independent ACGO, and machine that needs to be configured with independent ACGO
without three-way valve. Independent ACGO with three-way valve applies to the machine that
needs to be configured with ACGO with three way valve.
5.16 Module Rack
The setting of module rack provides two options: Open and Close.
When module rack is set to Open, the machine can support to identify module plugging and
monitor the fan speed of module rack.
When module rack is set to Close, the machine cannot support to identify module plugging or
monitor the fan speed of module rack.
5.17 AG Module
The setting of AG module provides two options: Open and Close.
When AG module is set to Open, the machine can identify the external AG module.
When AG module is set to Close, the machine cannot identify the external AG module.
5.18 BIS Module
The setting of BIS module provides two options: Open and Close.
When BIS module is set to Open, the machine can identify the BIS module.
When BIS module is set to Close, the machine cannot identify the BIS module.
5.19 CO2
Module
The setting of CO2 module provides two options: Open and Close.
When CO2 module is set to Open, the machine can identify the external CO2 module.
When CO2 module is set to Close, the machine cannot identify the external CO2 module.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...