8-90
8.6.1.4 Adjusting the Regulator of the Pipeline Gas Supply Inlet
Turn the regulator clockwise by using an M3 inner hexagon wrench to increase the pressure or
anticlockwise to decrease the pressure. See the figure below. Each time after you adjust the
pressure, release the pressure inside the pipeline gas supply inlet by using the calibration hose of
the regulator (PN0453-00-1216). Then turn on the pipeline gas source gain. Observe the adjusted
pressure on the test pressure gauge. Adjust the pressure to 200 kPa.
8.7 Hardware and Electrical Faults
Failure
Description
Possible Cause
Judgment Method
Recommended
Action
The
anaesthesia
machine
cannot be
powered on.
The system
switch is in poor
condition.
According to the startup signal flow,
remove the interconnection connector
between the system switch cable and
the airway cable, and short-circuit the
pins of the airway cable. If the
anaesthesia machine can be started,
the system switch is in poor condition.
If the anaesthesia machine cannot be
started, proceed with the analysis.
Replace the
system switch.
The cable is in
poor condition.
Short-circuit the J12 pins of the
mother board by using a metal tool. If
the anaesthesia machine can be
started, the airway cable or system
switch cable is in poor condition.
Replace the
cable.
The power board
is faulty.
If the anaesthesia machine still cannot
be started after the preceding two tests
are finished, the power board is
faulty.
Replace the
power board.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...