7-58
4. Check the airway and make sure that there are no occlusions or leaks.
Vent the sampling tubing to the air and check whether the current rate and set rate
are approximately the same. If the deviation is great, it indicates that there is an
occlusion in the tubing. Check the tubing for an occlusion.
Block the gas inlet of the sampling line. The current rate should drop rapidly and
the message of airway occlusion is prompted. Otherwise, it means that the airway
leaks. Check the airway for leakage.
5. Connect the test system as follows.
6. Open the relief valve and vent a certain standard gas or gas mixture and make sure that
there is an excess gas flow through the T-shape connector to air.
7. In
the
[O2
Module Cal.
] menu, the measured gas concentration and flow are displayed.
If the difference between the measured gas concentration and the actual one is
tolerable, a calibration is not needed.
If the difference is great, a calibration should be performed.
7. Select
[
Calibrate
] to start a calibration.
8. Enter the vented gas concentration and set other gases’ concentration to 0.
9. After a successful calibration, the screen shows [
Calibration Completed!
]. Otherwise,
the message [
Calibration Failure! Please try again.
] is displayed. In this case, you
need to do the calibration again
.
7.3.10.3 Commonly-encountered Problems and Recommended Actions
Failure description
Possible cause
Recommended action
Calibration menu is
inaccessible.
The AG module is not fully warmed up
or not in measure mode.
Wait for the AG module to be
fully warmed up and then
access the calibration menu.
Calibration is not
completed.
1. The module is damaged.
2. The difference between the set O2
calibration concentration and the
selected standard O2 concentration is
too great.
1. Return the module to factory
for repair.
2. The difference between the
standard gas concentration and
the set calibration concentration
cannot exceed 25% of the
standard gas concentration.
AG module
Gas cylinder
Relief valve
Tubing
Open to the air
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...