8-46
10. Switch off the O2 supply and keep the reading on the test pressure gauge remain
non-decreased. If the "O2 Supply Failure" alarm is generated in 10s, the pressure switch
of the O2 inlet assembly is damaged. Handle this issue according to the fault locating
table.
11. Adjust the O2 flow regulator until the O2 flow reaches 0.5 L/min, to gradually decrease
the reading on the test pressure gauge to 0.2 MPa (2 bar).
12. Switch off the O2 supply and keep the reading on the test pressure gauge remain
non-decreased. If the "O2 Supply Failure" alarm is generated in 10s, the pressure switch
of the integrated airway of the exhalation valve assembly is damaged. Handle this issue
according to the fault locating table.
13. Adjust the O2 flow regulator until the O2 flow reaches 0.3 L/min, to gradually decrease
the reading on the test pressure gauge to 0.15 MPa (1.5 bar).
14. Switch off the O2 supply and keep the reading on the test pressure gauge remain
non-decreased. If the "O2 Supply Failure" alarm is not generated in 10s, the pressure
switch of the O2 inlet assembly is damaged. Handle this issue according to the fault
locating table.
15. Adjust the O2 flow regulator until the O2 flow reaches 0.3 L/min, to gradually decrease
the reading on the test pressure gauge to 0.05 MPa (0.5 bar).
16. Switch off the O2 supply and keep the reading on the test pressure gauge remain
non-decreased. If the "O2 Supply Failure" alarm is not generated in 10s, the pressure
switch of the integrated airway of the exhalation valve assembly is damaged. Handle
this issue according to the fault locating table.
8.4.2.3 Adjusting the Pressure Switch
Adjust the O2 pressure switch and drive gas pressure switch.
Adjust the O2 pressure switch using a slot-head screwdriver. Screw a little bit each time, for
example, 30 degree. Note that the alarm limit is decreased when you rotate the pressure switch
clockwise and the alarm limit is increased when you rotate the pressure switch anticlockwise. Test
the component each time you adjust the pressure. Repeat the operations until the pressure switch is
correctly adjusted and the pressure is remained within the specified range (nominal value 220 kPa
+/- 10 kPa).
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...