2-63
The AGSS transfer system is a clear tube with 30 mm purple conical connectors at both ends, a
female 30 mm conical connector as the inlet and a male 30 mm conical connector as the outlet.
The transfer system is connected to the receiving system through the male 30 mm conical
connector. The following figure shows the AGSS structure and the connections between the AGSS
transfer system, receiving system, and disposal system.
2.2.8 Breathing System Heater
The breathing system heater of this anesthesia machine provides dual overheat protection for
software and hardware. The heater can change its operating mode automatically according to the
ambient temperature.
2.2.9 Ventilator Pneumatic - O2 Drive Gas
Oxygen or air, switching by the drive gas switch, is the drive gas for the ventilator. In addition to
the flow meter block, a high pressure regulator reduces the supply pressure of the drive gas for the
ventilator to 200 kPa (29 psi).
The drive pressure regulator is placed in front of the proportional valve that generates the drive
gas flow during the inspiratory phase. This flow is injected into the space between the bellows and
the bellows housing.
Main part of transfer and
receiving system
Hose of receiving system
To the waste gas exhaust port of the
anesthesia machine
To the disposal system
Hose of transfer system
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...