6-21
6.6.4 Check the Sensor Zero Point
NOTE
The zero point A/D value of the airway pressure sensor and PEEP pressure sensor
should fall within the normal range of 7432 to 16206.
The zero point A/D value of the inspiratory flow sensor should fall within the
normal range of 554 to 26457.
The zero point A/D value of the expiratory flow sensor should fall within the
normal range of 1392 to 19306.
The zero point A/D value of the built-in ventilator flow sensor should fall within the
normal range of 1699 to 16768.
If the zero point of the pressure sensor has an error, in ventilation status, the
baseline of the Paw waveform is not at the zero point and a great deviation occurs
between pressure control and measurement.
If the zero point of the inspiratory/expiratory flow sensor has an error, in
ventilation status, the baseline of the flow waveform is not at the zero point and a
great deviation occurs between TV control and measurement.
If the zero point A/D value of any sensor is outside of the normal range, it cannot
be corrected. The monitor board must be replaced.
Checking the sensor zero point is to check if the zero points of all the flow sensors and pressure
sensors inside the machine are within the normal range so as to determine if replacing the monitor
board is necessary.
To check the sensor zero point:
1. Turn off all fresh gases and position the Y piece connector in the patient circuit to the
air.
2. Make sure that the system is Standby. Select the [Maintenance] shortcut key
→
[
Factory Maintenance >>
]
→
[
Diagnostic Test >>
]
→
[
Display A/D Channels >>
]
→
[
Ventilator >>
] to access the [
Display A/D Channels—Ventilator
] menu.
3. Make sure that the actual measured value of each sensor is “0” (zero). Record the zero
point A/D value of each sensor and judge if the zero point falls with the normal range. If
not, replace the monitor board.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...