5-14
For the anesthesia machine configured with the function of Drive Gas Auto Switch, there will be a
[
Drive Gas Valve
] button added under the [
PPEP Valve D/A
] button in this menu.
After setting relevant items, check the A/D value and actual value in the valve test table on the
right to judge whether the test item is accurately controlled. The A/D value is significant for zero
verification. When no air flows through the sensor (for example, obstruct the air source, stop the
fresh air flow, or enter standby mode), the displayed A/D value is the present zero value. When
there is no offset at zero, the actual value shall be 0.
However, if offset occurs to the sensor zero point, the actual value displayed is not always 0. In
this case, zero calibration is required to adjust the measured value to 0. The ranges of zero point
AD value of each sensor are as follows:
Name of sensors
The ranges of zero point AD
value
internal flow sensor
1699-16768
inspiratory flow sensor
554-26457
expiratory flow sensor
1392-19306
airway pressure sensor
7432-16206
PEEP pressure sensor
7432-16206
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...