8-87
8. Turn on the pipeline air source and record the reading on the test pressure gauge. If the
difference between the readings of the test pressure gauge and pipeline air pressure
gauge exceeds 0.1 MPa (1.0 bar), the pipeline air pressure gauge is damaged. Handle
this problem according to the fault identification table.
9. Reconnect PU tube 41 to the pressure gauge.
8.6.1.2 Testing the Pressure Switch
Check that the pressure switches of the O2 supply inlet and expiratory valve comply with the
regulations by using the following tools:
1 MPa (1.0 bar) test pressure gauge (quantity: 1)
3106-04-06 adapter connector (quantity: 1)
3106-06-08 adapter connector (quantity: 1)
3140-08-00 Y-shape tube (quantity: 1)
PU tube (8x200) (quantity: 2)
PU tube (6x200) (quantity: 1)
PU tube (4x200) (quantity: 1)
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...