7-38
Failure
Description
Possible Cause
Recommended Action
[00008000] is
displayed.
The ACGO switch is in ON
position.
1. Check whether the ACGO switch is
in OFF position.
2. Check the monitoring status of the
ACGO switch on the interface of the
anesthesia machine.
[00010000] is
displayed.
The valve resolution is
insufficient.
1. Check whether the gas supply
volume is sufficient for the whole
calibration process.
2. Check whether the calibration
device works properly.
3. Replace the PEEP valve.
[00020000] is
displayed.
The maximum value cannot be
found.
1. Use the valve diagnosis tool to
diagnose: (1) Turn on the PEEP valve
in 4000 DA and check that the pressure
measured by the VT reaches 90 cm
H2O or higher.
(2) Turn off the PEEP valve. Open up
the PEEP valve gradually in DA. The
AD value measured by the flow
pressure sensor increases gradually.
If the above two conditions are met, a
re-calibration is recommended.
2. Replace the PEEP valve.
[00040000] is
displayed.
The minimum value cannot be
found.
1. Use the valve diagnosis tool to
diagnose: (1) Open up the PEEP valve
from 0 to 2000 DA. If the AD value
measured by the inspiratory pressure
sensor increases gradually, a
re-calibration is recommended.
2. Replace the PEEP valve.
[00080000] is
displayed.
The zero point of the calibration
device exceeds the standard.
1. Zero the calibration device.
[00100000] is
displayed.
The small pressure calibration is
not monotonic.
1. Check the connection between the
calibration device and the sampling
line.
2. Replace the PEEP valve assembly.
Содержание WATO EX-55Pro
Страница 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 34: ...2 20 South African standard ...
Страница 48: ...2 34 VPM ...
Страница 139: ...4 39 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 142: ...4 42 ...
Страница 144: ...4 44 b Gas Settings Press the Setup button select Setting ENTER Gas Settings MODIFY Gas Type O2 c Select BACK BACK BACK ...
Страница 196: ...6 26 FOR YOUR NOTES ...
Страница 213: ...7 17 2 Zero the sensor without connecting any tubes 3 User Setup Gas Type O2 4 User Setup Measure Mode BTPS ...
Страница 288: ...8 34 Test fixture of the vaporizer support Anesthesia machine calibration device 1 MPa 10 bar test pressure gauge ...
Страница 356: ...8 102 FOR YOUR NOTES ...
Страница 396: ...9 40 2 Remove the two ends of the breathing tubes from the inspiration expiration connectors on the breathing system ...
Страница 438: ...9 82 FOR YOUR NOTES ...
Страница 440: ...10 2 10 3 Diagrams and Tables 24 17 16 7 6 5 3 2 18 20 25 1 4 23 21 22 15 14 13 12 11 10 9 8 26 19 ...
Страница 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Страница 449: ...10 11 ...
Страница 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Страница 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Страница 462: ...10 24 4 3 1 2 ...
Страница 473: ......
Страница 474: ...PN 046 009688 00 4 0 ...